Contact an Account ManagerDetermination of remoxipride in plasma and urine by HPLC
Summary
Type: HPLC
Application #: 104412
Catalog#: 720149.46 - MN HPLC Column, Nucleosil C18, 120A, 3um, 4.6x100mm
Detection: UV, 206nm
Conditions: eluent: acetonitrile - phosphate buffer pH 2 (25:75, v/v) with 0.2mM N,N-dimethyl-N-nonylamine
flow rate 1.3ml/min
Authors: Nilsson LB
Source: J Chromatogr 526 (1990) 139 - 150
Details
Legend: Fig. 2 from 1st paper. Human plasma chromatogram 6h after a 100mg oral dose of
remoxipride (R), 2.49µM, I.S. internal standard, 4µM.